Series 661 Force Transducer: Specifications - MTS Systems 318 Series Product Information

Load unit
Table of Contents

Advertisement

Series 661 Force Transducer: Specifications

M
L
ODEL
661.11-01
250 N
(50 lbf)
661.11-02
500 N
(100 lbf)
661.18-01
1 kN
(220 lbf)
661.18-02
2.5 kN
(550 lbf)
661.19-01
5 kN
(550 lbf)
661.19-02
10 kN
(2.2 kip)
661.19-03
15 kN
(3.3 kip)
661.19-04
25 kN
(5.5 kip)
24
Introduction
The force transducer used with this system is a Series 661 Force Transducer. The
following are the specifications for the force transducers.
P
ARAMETER
Maximum excitation voltage
Bridge resistance
Maximum crosstalk
Hysteresis
Nonlinearity
Temperature
Usable range
Compensated range
Sensitivity
Output
Connector
C
OAD
APACITY
T
HREAD
M6 x 1.0 mm x 6.3 mm
(1/4 - 28 UNF x 0.25 in
M6 x 1.0 mm x 6.3 mm
(1/4 - 28 UNF x 0.5 in
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
M12 x 1.25 mm x 25.4 mm
(1/2 - 20 UNF x 1.0 in)
S
PECIFICATION
15 V DC
350 ¾
1.0% of full scale torsional to load
0.08% of full scale (250 N–2.5 kN)
0.05% of full scale (5 kN–50 kN)
0.15% of full scale (100 kN–500 kN)
0.20% of full scale (1000 kN)
0.08% of full scale
0.15% of full scale for Models 661.22/.23/.31
0.004% of reading/°C (0.002%/°F)
-54°C (-65°F) to +121°C (+250°F)
+21°C (+70°F) to +77°C (+170°F)
0.0036% of full scale/°C
(0.0020% of full scale/°F)
2 mV/V at full-scale load
PT02ER-10-6P
IZE *
S
W
EIGHT
0.45 kg
(1 lb)
0.45 kg
(1 lb)
2.27 kg
(5 lb)
2.27 kg
(5 lb)
3.07 kg
(6.75 lb)
3.07 kg
(6.75 lb)
3.07 kg
(6.75 lb)
3.07 kg
(6.75 lb)
318 Load Unit

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

318.10318.25318.50

Table of Contents