Process Calibration; Start The Process Calibration - Swann AMI Phosphate-II Operator's Manual

Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
6.6.
6.6.1
A-96.250.641 / 290121

Process Calibration

The process calibration is based on a comparison measurement of
the Instrument with an external high precision instrument. A process
calibration can only be performed if the instrument has finished at
least one valid measurement.
Consider that the accuracy of the process calibration depends on the
following factors:
 The time between taking a sample and its measurement
should be as short as possible to avoid contamination of the
sample with unknown substances.
 Sampling should be done as close as possible on the location
of the measurement.
 The sample container must be cleaned well.
 The accuracy of the measuring instrument.
Under certain circumstances a calibration is more precise than a
process calibration.

Start the Process Calibration

Prozesscal.
Current Value
2.29 ppm
Slope
Process value
2.35 ppm
<Enter> to save
Prozesscal.
Current Value
2.35 ppm
Slope
Process value
2.35 ppm
Calibration Successful
<Enter> to continue
Possible error messages:
 Slope error
 Current value invalid
If a slope error occurs, the slope value is not saved. If one of these
errors occurs proceed as follows: Repeat the process calibration and
avoid possible sources of error.
1 Navigate to menu 3.2
3.2.5
<Maintenance/Process cal.
2 Press [Enter]
0.96
3 Enter the measured comparison
process value.
4 Press [Enter] to save the value.
3.2.5
 The slope is recalculated and
displayed.
0.98
43

Advertisement

Table of Contents
loading

Table of Contents