AMI Phosphate-II
Default Values
11.
Operation:
Sensors:
Alarm Relay
Relay 1 and 2
Input
Logger:
Installation:
Sensors
Signal Output
1 and 2
Alarm Relay:
Relay1 /2
A-96.250.641 / 290121
Default Values
Filter Time Const.:........................................................................ 30 s
Hold after Cal.:........................................................................... 300 s
.........................................................................same as in Installation
.........................................................................same as in Installation
.........................................................................same as in Installation
Logger Interval:.............................................................. Event-driven
Clear Logger: .................................................................................. no
Ref. Verification:......................................................................... 0.235
Phosphate as:.............................................................................. PO4
Standard PO4: ....................................................................... 1.0 ppm
Meas. Interval: ......................................................................... 10 min
Channels:.......................................................................................... 1
Channel selection: .................................................................. internal
Cleaning:............................................................................ 2 solutions
Parameter: ...................................................................... Phosphate 1
Current loop: ........................................................................ 4–20 mA
Function: .................................................................................... linear
Scaling: Range low: ............................................................. 0.00 ppm
Scaling: Range high:.......................................................... 10.00 ppm
Alarm Phosphate1:
Alarm high:......................................................................... 10.00 ppm
Alarm low: ............................................................................ 0.00 ppm
Hysteresis: ........................................................................... 0.10 ppm
Delay:............................................................................................. 5 s
Sample Flow: Flow Alarm: .............................................................yes
Sample Flow: Alarm High: ...................................................... 500 B/s
Sample Flow: Alarm Low: ......................................................... 10 B/s
Case temp. high:........................................................................ 65 °C
Case temp. low: ........................................................................... 0 °C
Function: .............................................................................limit upper
Parameter: ...................................................................... Phosphate 1
Setpoint:............................................................................... 10.0 ppm
Hysteresis: ........................................................................... 0.10 ppm
Delay:........................................................................................... 30 s
83