Maintenance
6.8.
Material
Procedure
A-96.250.641 / 290121

Cleaning the Photometer

Clean the photometer after indication by alarm (E020, FOME dirty).
Switch off the instrument according to instructions in
tion for Maintenance, p.
Small brush.
B
C
1 Close the flow regulating valve [A].
2 Unscrew the cover [B] from the photometer [C].
D
C
3 Clean the Photometer with a small brush [D].
4 Screw the cover to the photometer.
5 Open the flow regulating valve.
37.
A
Flow regulating valve
B
Photometer cover
C
Photometer
A
D
Brush