AMI Phosphate-II
Troubleshooting
Error
Description
E019
FOME not connected
E020
FOME dirty
E022
Reagent empty
E023
Cleaning Solution
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEprom Frontend
E031
Calibration Recout
E032
Wrong Frontend
E049
Power-on
E050
Power-down
E065
Reagents low
E067
Cleaning Solution
A-96.250.641 / 290121
Corrective action
– shut off power
– check wiring
– clean photometer, see
Photometer, p. 45
– refill reagents, see
p. 38
– refill Cleaning Solution
– See If Fault Yes is programmed in Menu
p. 79
– Hardware failure, call service
– Check wiring on signal outputs 1 and 2
– Hardware failure, call service
– Hardware failure, call service
– Hardware failure, call service
– none, normal status
– none, normal status
The decreasing number next to the triangle
in the upper status line on the display, indicates
the remaining reagent in percent. Refill reagent
on time, see
Refill or replace Reagents, p. 38
Only AMI Phosphate II with Cleaning Module. A
triangle without number
cleaning solution containers are empty.
– Refill cleaning solution
Cleaning the
Refill or replace Reagents,
indicates that the
5.3.4,