AMI Phosphate-II
Troubleshooting
7.2.
58
Opening the peristaltic pump housing
For some electrical connections (e.g. when replacing suction lanc-
es), it is necessary to open the housing of the peristaltic pump. To do
this, proceed as follows:
1 Switch off the analyzer according to
nance, p.
37.
2 Remove the protection cap and all pump tubes as described in
Dismount pump tubes, p.
3 Unscrew the 4 screws of the peristaltic pump housing and re-
move the cover.
4 Disconnect the motor connector [A].
5 Feed the cable into the housing through one of the PG7 cable
glands.
6 Connect the cable to the terminal block of the peristaltic pump ac-
cording to
Connection Diagram, p.
7 Reassemble in reverse order.
Stop of Operation for Mainte-
52.
A Motor connector
A
21.
A-96.250.641 / 290121