AMI Phosphate-II
Program List and Explanations
5.2.1.43
5.2.1.43.10
5.2.1.43.20
5.2.1.43.3
5.2.1.43.3
5.2.1.43.3
5.3 Relay Contacts
5.3.1
A-96.250.641 / 290121
Control upwards or downwards
Setpoint: User defined precess value for the selected parameter.
P-Band: Range below (upwards control) or above (downwards con-
trol) the set-point, within which the dosing intensity is reduced from
100% to 0% to reach the set-point without overshooting.
Control Parameters: Process value Phosphate 1
Setpoint:
Range: 0.00 –10.00 ppm
P-Band:
Range: 0.00 –10.00 ppm
Reset time: The reset time is the time till the step response of a sin-
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9'000 sec
Derivative time: The derivative time is the time till the ramp response
of a single P-controller will reach the same value as it will be sudden-
ly reached by a D-controller.
Range: 0–9'000 sec
Control time out: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value does
not come closer to the setpoint, the dosing process will be stopped
for safety reasons.
Range: 0–720 min
Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
Power loss
Detection of system faults like defective sensors or electronic
parts
High case temperature
Lack of reagents
Process values out of programmed ranges.
Program alarm levels, hysteresis values and delay times for the fol-
lowing parameters:
Phosphate, Sample Flow or Case Temperature
The alarm values of Phosphate, Sample Flow and Case Tempera-
ture can be programmed in menu as well
(5.3.1.3, p.
76).
75