AMI Phosphate-II
Program List and Explanations
5.2 Signal Outputs
5.2.1 and 5.2.2
5.2.1.1
5.2.1.2
5.2.1.3
As process
values
72
Note: The navigation in the menu <Signal Output 1> and <Signal
Output 2> is equal. For reason of simplicity only the menu
numbers of Signal Output 1 are used in the following.
Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
Parameter: Assign one of the process values to the signal output.
Available values: Phosphate, Sample Flow.
Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
Function: Define if the signal output is used to transmit a process val-
ue or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
As process values, p. 72
Control upwards or control downwards for controllers.
See
As control output, p. 74
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
B
10
12
0 / 4
0.0
0.1
0.2
A
linear
B
bilinear
A
0.3
0.4
0.5
X Measured value
X
A-96.250.641 / 290121