AMI Phosphate-II
Program List and Explanations
5.3.2.54
5.3.2.6
5.3.2.7
5.3.2.1
5.3.3.1
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
A-96.250.641 / 290121
Delay: during run time plus the delay time the signal and control out-
puts are held in the operating mode programmed below.
Range: 0–6'000 Sec
Signal Outputs: select the behavior of the signal outputs when the re-
lay closes. Available values: cont., hold, off
Output/Control: select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
Function = Fieldbus
The relay will be switched via the Profibus input. No further parame-
ters are needed.
Function = End of Batch
This function is only available on relay 2. It is used to communicate
with canal switching instruments from third-party suppliers. If End of
Batch is selected, no further selection is possible.
Input: The functions of the relays and signal outputs can be defined
depending on the position of the input contact, i.e. no function,
closed or open.
Active: Define when the input should be active:
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Continuous:
Signal outputs continue to issue the measured
value.
Signal outputs issue the last valid measured value.
Hold:
Measurement is interrupted. Errors, except fatal er-
rors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Continuous:
Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.
79