Refill Or Replace Reagents - Swann AMI Phosphate-II Operator's Manual

Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
6.3.
Canister setup
Reagent
consumption
38

Refill or replace Reagents

The liquid level in container 2 is monitored. The following message is
displayed:
Container almost
empty
Container empty
CAUTION
Pollution of reagents
Refilling reagents without flushing the containers may lead to pol-
lution of the reagents.
 Before refilling the reagents, rinse the container with deminer-
alized water.
A
B
C
D
E
F
G
Depending on the measuring interval, a 2-litre reagent container is
sufficient for an operating period of between 1 and 3 months.
Maintenance E065 - Reagents low and the
remaining reagent volume in% (starting at
17% = 340ml). See
Error E022 - Reagent empty
Operation, p. 32
A
Suction lance without
level detector
(container 1)
B
Suction lance with
level detector
(container 2)
C
Level detector
D
2 L mark
E
Reagent container 1
F
Reagent container 2
G
Holder
A-96.250.641 / 290121