AMI Phosphate-II
Troubleshooting
Error
Description
E001
Phos. 1 Alarm high
E002
Phos. 1 Alarm low
E003
Phos. 2 Alarm high
E004
Phos. 2 Alarm low
E005
Absorbance too high
E009
Sample Flow high
E010
Sample Flow low
E013
Case Temp. high
E014
Case Temp. low
E015
Valve defective
E017
Control Timeout
E018
Reagent Pump
56
Corrective action
– check process
– check programmed value
– check process
– check programmed value
– check process
– check programmed value
– check process
– check programmed value
– check process
– check sample flow
– check programmed value
– adjust sample flow with the flow regulating
valve
– establish sample flow
– clean instrument
– check programmed value
– check environment temperature
– check programmed value
– check environment temperature
– check programmed value
– check valve, see 2
– check control device or programming in
Installation, Relay contact, Relay 1/2
5.3.3, p. 77
– shut off power
– check wiring
5.3.1.1.1, p. 76
5.3.1.1.25, p. 76
5.3.1.1.1, p. 76
5.3.1.1.25, p. 76
5.3.1.3.2, p. 76
5.3.1.3.35, p. 76
5.3.1.4, p. 76
5.3.1.4, p. 76
49
5.3.2 and
A-96.250.641 / 290121