Swann AMI Phosphate-II Operator's Manual page 81

Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Program List and Explanations
5.5 Interface
5.5.1
5.5.20
5.5.30
5.5.40
5.5.1
5.5.21
5.5.31
5.5.41
5.5.1
5.5.1
A-96.250.641 / 290121
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Protocol: Profibus
Device address:
ID-Nr.:
Local operation:
Protocol: Modbus RTU
Device address:
Baud Rate:
Parity:
Protocol: USB-Stick:
Only visible if an USB interface is installed. No further settings are
possible.
Protocol: HART
Device address:
Range: 0–126
Range: Analyzer; Manufacturer; Multivariable
Range: Enabled, Disabled
Range: 0–126
Range: 1200–115 200 Baud
Range: none, even, odd
Range: 0–63
81