Electrical Requirements

NOTE:
GE Medical Systems requires a dedicated power and Ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-5-1
Vivid™ 3 Power Requirements
Electrical specifications for the Vivid™ 3 monitor and onboard peripherals are as follows:
Table 2-3
Voltage
100V AC
120V AC
220 - 240V AC
2-2-5-2
Inrush Current
Inrush current is not a factor for consideration, due to the inrush current limiting properties of the power
supplies.
Maximum power requirement = 1.2 KVa
100V AC: 8A
120V AC: 8A
220 - 240V AC: 4A
2-2-5-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
POWER OUTAGE MAY OCCUR.
CAUTION
The Vivid 3 requires a dedicated single branch circuit. To avoid circuit overload and possible
loss of critical care equipment, make sure you DO NOT have any other equipment operating on
the same circuit.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Electrical Requirements
Tolerances
Op. Current
±10%
±10%
±10%
Chapter 2 - Pre-Installation
V
IVID
Frequency
8A
50-60 Hz
8A
50-60 Hz
4A
50-60 Hz
™ 3 P
/V
™ 3 S
RO
IVID
ERVICE
M
ANUAL
2-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 3

Table of Contents