CFM Operation Tests
Table 4-24
Step
1
2
3
4
4-14
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
CFM Operation Tests (3S)
Task
Select a Cardiac preset and set the 2D Gain
to 40.
Activate CFM and set the following CFM
parameters:
Gain: Max possible without color noise
Frame Rate:2
LVR: MIN setting
CFM Artifacts Test
Place the 3S probe on the RMI1425A
Phantom (top view A).
CFM Alignment Test
Rotate the probe 90 degrees (top view B) to
see a circular cross-section of the vessel.
Section 4-4 - Image Testing: 2D/M/CFM/Doppler
V
™ 3 P
IVID
RO
Expected Result(s)
Ensure there are no flash strikes or other color artifacts.
Ensure that the color exactly overlaps the vessel.
/V
™ 3 S
M
IVID
ERVICE
ANUAL