GE Vivid 3 Pro Service Manual page 130

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
NOTE:
IMPORTANT INFORMATION ABOUT DATAFLOWS:
The system's default dataflow is Local Archive - Int. HD. In this configuration, the Patient List is in the
local archive and the images are stored on the internal hard drive.
In a dataflow configuration such as Int. HD - MOD 525 (DICOM), images are stored both on the internal
hard drive and on media (MOD). The name DICOM is shown in parenthesis, signifying that on the MOD,
the images are in DICOM format.
No Archive is a dataflow that has no output source, for use in emergencies.
CAUTION: DO NOT change the name of an existing dataflow or attempt to add new services to
a dataflow! In the event that any changes or additions are required, contact the On-line Center.
3-7-6-2
Validating EchoPAC or Image Vault Operation
1) Press the Archive button on the control console. The Patient List window is displayed, as shown in
Figure 3-44
2) Select the Remote Archive-Remote Storage option from the dropdown list at the bottom of the
window. If the system will not accept the dataflow, or displays an error message, there is an error
in dataflow communication.
3) Begin an exam for a new patient and store some images. The storage process will take longer as
each loop is now stored in both raw data and DICOM format.
4) Begin an exam for a new patient. Once the second patient is active, the first patient's exam is closed
and sent to the spooler to be sent over the network.
5) Press <Ctrl+S> to view the status of the spooler, as described on page 3-63.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
on page 3-62.
Chapter 3 - Installation
V
™ 3 P
/V
™ 3 S
IVID
RO
IVID
M
ERVICE
ANUAL
3-67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 3

Table of Contents