RFT Diagnostic Tests (for RFT Configuration)

The RFT diagnostic tests include the tests listed below:
Board Test: General RFT board test which activates the remaining tests sequentially. Each of
these tests can also be performed independently.
Self Test 1: Short generic board test, generated in a local DSP.
Self Test 2: High level board test, generated in a local DSP. Self Test 2 lasts six minutes.
Note:
Before performing the RFT test, perform the IMP and VME Bus tests, as described in the
Port (IMP) Diagnostic Tests (for RFT Configuration)
(VME) Diagnostic Tests (for RFT Configuration)
performed without first performing these tests, and errors are identified, you must perform the
VME Bus test, and then the IMP test, to ensure that the RFT board is the source of the error.
1) Access the FE Diagnostics dialog box, as described in the
Options
2) In the FE Diagnostics dialog box, trackball to the RFT tab and press Select,
OR
Trackball to the RFT button in the Data Flow Map and press Select.
The RFT tab is displayed, as shown below:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
section, on page 7-6.
Figure 7-9 RFT Tab - FE Diagnostics Dialog Box
Chapter 7 - Diagnostics/Troubleshooting
V
™ 3 P
IVID
RO
section, on page 7-13 and in the
section, on page 7-15. If the RFT test is
Accessing the Front End Diagnostic
/V
™ 3 S
M
IVID
ERVICE
ANUAL
Image
VME Bus
Data Flow Map
Test Description Area
Test Area
Comment Area
Tabs