System Voltage Confirmation

3-4-4-1
System Voltage Settings
Verify that the scanner is set to the correct voltage.
The Voltage settings for the Vivid™ 3 Scanner are found on a label to the right of the Power switch and
External I/O, on the rear of the system.
Figure 3-13 Rating Plate Example
WARNING: CONNECTING A VIVID
MOST LIKELY DESTROY THE SCANNER.
3-4-4-2
Confirming System Voltage Configuration
1) Turn ON the system.
2) In regular 2D Scanning Mode, press Config.
3) From the System Configuration dialog box, click the Technical Support tab.
4.) Make sure the frequency and voltage ranges are set up correctly and that the appropriate settings
are displayed in the Power Supply Frequency and Nominal Voltage fields (lower right side of the
dialog box - see
5) Click OK.
Note:
If the voltage is not set correctly, contact an authorized GE Service Representative.