GE Vivid 3 Pro Service Manual page 268

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
7) Proceed as follows:
If no problems were identified, proceed to the FB tests, as described in the
Assembly (FB) Diagnostic Tests
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
engineer for assistance.
8) When finished, trackball to the Cancel button and press Select. The FE Diagnostics dialog box is
re-displayed.
7-24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
section, on page 6-14, and then repeat the test.
(Theory). If the problem persists, contact your local GE Medical Systems field
Section 7-3 - Performing Front End (FE) Diagnostics
V
™ 3 P
IVID
section, on page 7-25.
/V
™ 3 S
M
RO
IVID
ERVICE
ANUAL
Front Board
Chapter 5 - Components

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 3

Table of Contents