Video OUT Test
Table 4-8
Step
1
2
3
4
5
4-3-2-3
B/W Printer Tests
Table 4-9
Step
1
2
3
4-3-2-4
Color Printer Tests
Table 4-10
Step
1
2
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Video OUT Test
Task
Activate a CineLoop or Freeze a scanned image.
Connect the monitor to the composite VIDEO OUT 1
connector and check the image.
Connect the monitor to the composite VIDEO OUT 2
connector and check the image.
Connect the monitor to the composite VIDEO B/W
connector and check the image.
See
B/W Printer
Tests, below.
Connect the monitor to the Color Printer RGB
connectors and check the image.
See
Color Printer
Tests, below.
B/W Printer Test
Task
Scan a phantom or suitable interface in 2D-Mode.
Press the Freeze button.
Press the Print B button.
Color Printer Test
Task
Scan a blood vessels in 2D-Mode or a phantom
simulating a similar CFM effect, then activate CFM to
see color.
Press the Freeze button.
Press the Print A button.
Chapter 4 - Functional Checks
V
™ 3 P
/V
™ 3 S
IVID
RO
IVID
Expected Result(s)
The image is displayed correctly.
The image is displayed correctly.
The image is displayed correctly.
The image is displayed correctly.
Expected Result(s)
Evaluate the print quality. Adjust the brightness and/or
contrast of the B/W printer, if required.
Expected Result(s)
Evaluate the color print quality. Adjust the brightness
and/or contrast of the Color printer, if required.