Rf And Tissue Processor Board (Rft) - GE Vivid 3 Pro Service Manual

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
5-4-7

RF and Tissue Processor Board (RFT)

The RF and Tissue Processor Board (RFT) receives data from the BF1 Board in the Front End. Both
types of data samples (RF_MODE) are communicated to the RFT Board from the Front End Controller
Board, via the Front End Bus.
5-4-7-1
Control
The RFT Board has a local DSP with an external EPROM. The DSP performs the following control tasks:
Handles communication with the BE over the VME Bus and the PCI Bus.
Updates filter coefficients and other parameters in the data path.
When output events are received from the FEC, the RFT Board sets up the pipe link to output
control, for data tagging and transfer.
During transfer of time motion data, the RFT Board creates a "time slot" data transfer, which
replaces the output event base transfer.
RFT Board
5-24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Figure 5-17 RFT Processor Board
Section 5-1 - Overview
V
™ 3 P
/V
™ 3 S
IVID
RO
IVID
To
Image Port
Board
M
ERVICE