Image Testing: 2D/M/Cfm/Doppler; Probe Image Quality Tests - GE Vivid 3 Pro Service Manual

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
Section 4-4

Image Testing: 2D/M/CFM/Doppler

NOTE:
To perform these tests thoroughly, it is recommended to use three types of soft tissue phantoms: RMI
403GS, RMI 404G and RMI 1425A, or an equivalent surface that can simulate the same soft tissue
effects.
4-4-1
3S Probe Image Quality Tests
4-4-1-1
2D Noise Uniformity Test
Table 4-18
Step
1
2
4-4-1-2
2D Performance Test
Table 4-19
Step
1
2
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
2D Noise Uniformity Test (3S)
Task
With a 3S probe in the air, select a
Cardiac preset.
Set the following parameters:
2D Gain = 50
Depth = 7cm
Dynamic Range = 80dB
2D Performance Test (3S)
Task
Use the Standard Imaging Phantom
RMI403GS.
Select a Cardiac preset.
Set the TGC so that the image is uniform.
Chapter 4 - Functional Checks
V
™ 3 P
IVID
RO
Expected Result(s)
Ensure that there are no white circular lines or other patterns in
the noise.
Expected Result(s)
/V
™ 3 S
M
IVID
ERVICE
ANUAL
4-11