Accessing the Front End Diagnostic Options

In 2D-Mode, press <Ctrl+F11> on the alphanumeric keyboard. The Diagnostic menu is displayed,
as shown in
NOTE:
It is recommended to commence Front End Diagnostic tests using the Auto Sequence option, as
described below. To perform tests using the manual options, see
7-3-1-1
Auto Sequence
1.) In the Diagnostic menu, trackball to the Auto Sequence button and press Select. The Automatic
FE Testing dialog box is displayed, as shown below:
2.) Select the required number of Loops and activate the Enable checkbox.
3.) Under On Failure, activate either the Stop or Continue checkbox, as required. Stopping on failure
enables you to review a failed test, the moment it fails. Alternatively, you may prefer testing to
continue on failure and later you may review the report.
4.) Under Execute Sequence, trackball to the Start button and press Select.