7) Check the test result message and proceed as follows:
If one or more channels were identified as bad, trackball to the Save Report button and press
Select, then contact a GE Medical Systems Service Representative.
If no channels were identified as bad, mark the channels suspected as being bad as follows:
If no errors were identified, proceed with the next test.
If no errors were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
engineer for assistance.
8) Trackball to the Exit button and press Select. If required, repeat this test with a different pulser.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Trackball to each of the channels in the Display area and press select.
Trackball to the Mark bad channel checkbox and press Select.
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.
section, on page 6-14, and then repeat the test.
(Theory). If the problem persists, contact your local GE Medical Systems field