GE Vivid 3 Pro Service Manual page 163

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
4-4-4-5
CFM Operation Tests
Table 4-37
Step
1
2
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
CFM Operation Tests (739L)
Task
Select a Cardiac preset and set the 2D Gain
to 40.
Activate CFM and set the following CFM
parameters:
Gain: Max possible without color noise
Frame Rate:2
LVR: MIN setting
CFM Artifacts Test
Place the 739L probe on the RMI1425A
Phantom (top view A).
Chapter 4 - Functional Checks
V
™ 3 P
IVID
RO
Expected Result(s)
Ensure there are no flash strikes or other color artifacts.
/V
™ 3 S
M
IVID
ERVICE
ANUAL
4-23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid 3

Table of Contents