Technical Support Tab

The System Configuration Technical Support tab is used when new hardware has been installed, or
when it necessary to enable new functionality. The appearance of the Technical Support tab differs in
accordance with the specific Vivid™ 4 BT03 model (RFI or RFT configuration), as described below.
RFI-configured systems
RFT-configured systems
Monitoring
Watchdog
MOD Fan
Use Front End
ILinq
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Figure 3-28 Technical Support Tab
- should always be enabled (this provides constant monitoring of critical
parameters to ensure safe operation of the system).
- normally should not be enabled. When enabled, if there is a system lock-up
caused by the Back End, this will turn OFF power at the Front End.
- should be enabled only on BEPs based on Pentium IV - this provides faster
operation of the MOD. Enabling this option in the wrong type of BEP may
cause damage to the MOD or MO media.
- normally should be enabled.
- should be enabled only on systems that use the iLinq option.