2300164-100, R
IRECTION
The FE Diagnostics dialog box includes the following options:
Table 7-2
Option
Data Flow
Map
Test
Description
Area
Test Area
Comment
Area
Tabs
H/W Report
Button
Current
Report
Button
Close Button
7-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
FE Diagnostics Dialog Box Options
Description
Displays a graphic representation of the Vivid™
3 data flow. The functionality of each button in
this area corresponds to that of the tab of the
same name in the lower portion of the dialog
box. Each button or tab enables you to select
and perform the FE diagnostic tests by
trackballing to the required button or tab and
pressing Select.
Displays a description of the selected diagnostic
test.
Displays the buttons that enable you to perform
diagnostic tests.
Displays instructions during the test, as well as
test status and result messages.
Enable you to select and perform the FE
diagnostic tests.
Displays hardware and software version
information, as described in
Chapter 1 -
Introduction.
Displays the saved error log, as described in the
Common Service Interface
Closes the FE Diagnostic dialog box and
re-displays the Diagnostic menu.
Section 7-3 - Performing Front End (FE) Diagnostics
V
™ 3 P
IVID
Each button has an LED that indicates the
following by its color:
• Green: The test has been completed and no
problems were identified.
• Red: Problems were identified during the
performance of the test.
• Yellow: The test is in progress.
Each button has an LED that indicates the
following by its color:
• Green: The test has been completed and no
problems were identified.
• Red: Problems were identified during the
performance of the test.
• Yellow: The test is in progress.
The following messages are displayed:
• Test status: For example, Ready for test or
Complete
• Instructions: During the course of a
diagnostic test, any instructions to the user
are displayed in the Comment area
• Test results: For example, Pass or Fail.
section, on page 7-84.
/V
™ 3 S
M
RO
IVID
ERVICE
ANUAL