7-3-9-6-3 LVPS Voltage Test
The LVPS test measures the LVPS voltages. The voltage range is displayed next to the measured
voltage. All the voltages can be measured with DVM on the front board test point.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the
Diagnostic Test Options
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
in
Figure 7-36
4) Trackball to the LV PS Voltage button and press Select. The LVPS voltage ranges are displayed,
as shown below:
5) Trackball to the Show button and press Select to display the voltages.
Note: For repeat readings, select Loop every and set the reading time cycle.
7-62
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
section, on page 7-47.
on page 7-60.
Figure 7-38 LVPS Voltage Ranges
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
Figure 7-34
on page 7-58.
™ 3 P
/V
™ 3 S
RO
IVID
ERVICE
Accessing the MUX
M
ANUAL