GE Vivid 3 Pro Service Manual page 308

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
7-3-9-6-5 Temp Test
The Temp test displays the temperature at the Front End crate, as well as the temperature of the AC
distribution box.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in the
Diagnostic Test Options
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
in
Figure 7-36
4) Trackball to the Temperature button and press Select. The temperatures are displayed, as shown
below:
5) Trackball to the Show button and press Select to display the temperatures.
Note: For repeat readings, select Loop every and set the reading time cycle.
7-64
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
section, on page 7-47.
on page 7-60.
Figure 7-40 Temperatures
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
Figure 7-34
on page 7-58.
™ 3 P
/V
™ 3 S
RO
IVID
ERVICE
Accessing the MUX
M
ANUAL