Other Constants (continued)
Flow Span Enter the multiplication factor (flow rate/volt) used in conjunction with
Flow_Offset to convert voltage input from flowmeter into actual sample flow
rate. If the internal mass flowmeter is installed, the value is factory set and
should only be changed if the flowmeter is replaced.
Flow Offset Enter the additive factor (flow rate) used in conjunction with Flow_Span to
convert voltage input from flowmeter into actual cell temperature. If the
internal mass flowmeter is installed, the value is factory set and should only
be changed if the flowmeter is replaced.
Dry_CF This parameter, dry cell frequency, is the nominal cell frequency determined
during factory calibration of the instrument and is given for informational
purposes only.
Cell_SN Analyzer cell serial number. This value should only be changed if a new cell
is installed.
MoistGen_SN Moisture generator serial number. This value should only be changed if a
new moisture generator is installed.
Display (Display Flags)
This menu item is used to select the system values to be continuously displayed on the first three
lines of the display screen. Select which line is to be altered by highlighting Line_X_Flag where X
= 1 ⇒ 3. Enter the flag value from the following list when the "New Value?" prompt is displayed.
Press
to save change and return to the Configure Menu; press
enter
noted, the same flags are used for the analog output and alarm flags (Chapter 5).
Display Flag Values
0 None
1 Held Sample Moisture During Equilibration and Cal * 1.00 ppm Vol H2O
6 Flow Rate in sccm (if internal flowmeter only)
9 Live Moisture During Equilibration and Cal
10 Cell pressure in psia
11 Cell Temperature in
12 Cell Frequency (Cell_Freq) in Hz
13 Delta Frequency (Delta_Freq) in Hz
14 Status of Valve and Alarm Relays
15 Period Timers & Cal_Time OR Cal_Timer
(See
on page 5-5:Timing)
note
16 Output Range (Channel 1)
17 Zero_Time OR Zero_Timer
* These labels can be altered by the user; refer to Labels section for more information.
°
C
to quit Except where
cancel
Default Display
(Values are for Example ONLY)
Blank
* 200 sccm Flow
* 2.000 Current H2O
* 14.7 psi Cell Press
* 60.1 Degree Cent
1200 hz Cell Freq
0.50 hz Delta Freq
** 010-Alr 00001010 Vlv
** 27-Ref 08:00-Cal
or 27-Ref 003599-Cal
** -2.5
2.5 Rcdr Range
NOT USED
Software Configuration 5-3