Equation 3
● Equation 3 can be used to estimate the length of the high flow tube that is required. For
example, if an Hplc method uses a flow of 1.0 ml/min, using "Equation 2" to achieve a
flow rate of 1 µl/min in the spraychip (low flow tube):
For the SFI the following is constant:
Using Equation 1:
If the inner diameter of your desired waste tube is 254 µm:
It is always better to start with a longer piece of tubing (at least 10 cm longer) so that if the low
NOTE
flow rate is too high, the length of the waste tube can be cut to provide a suitable flow rate.
78
Flow SplITTInG
|
VERITY® 1910 MS DETEcToR