10. Lay the assembly face down on a clean lint-free surface.
● The tube lens assembly is
attached to the vac-chip
flange with five vacuum
screws. The vac-chip is
located under this assembly.
11. Use a Phillips screwdriver
to partially unscrew the five
vacuum screws in the following
pattern: 1, 4, 2, 5, 3 (Figure 20).
Repeat until the screws are
loose.
5
4
Figure 20
Tube Lens Assembly and
Vacuum Screws
12. Gently lift the tube lens assembly away from the vac-chip flange.
13. Use a pair of tweezers to remove the vac-chip from the flange (Figure 22).
The vac‑chip can become stuck on the tube lens assembly. Check that it is not before
NOTICE
resting the assembly to one side.
Assembly
1
2
3
Figure 21
The vac‑chip™ and Tube Lens Assembly
Figure 22
How to Handle the vac‑chip™
vac-chip
Screws
Tube
Lens
VERITY® 1910 MS DETEcToR
vac-chip
vac-chip
Flange
51