Safety Interlock

The VERITY 1910 MS Detector has a safety interlock system to protect users from high
voltages. This system is operated by the door in the front cover. The door has two positions
as shown in Figure 19.
When the door is closed, the safety interlock is off and high voltages can be applied. When
the door is fully closed it should be pushed gently forward until the catch engages with the
front cover. The catch is designed to prevent accidental activation of the safety interlock.
When the door is opened, the safety interlock is engaged and high voltages are disabled. The
door will click into position when the door reaches 90° and will remain fixed in this orientation
until pushed close.
Door closed
Figure 19
Front Cover Door Positions on the
VERITY® 1910 MS Detector
Door open
VERITY® 1910 MS DETEcToR