4. Gently slide the SFI into the vac‑chip assembly until the PEEK screw reaches the threaded
section of the vac‑chip assembly.
NOTICE
5. Screw the PEEK collar into the assembly until finger tight.
NOTICE
6. connect the high voltage connector from the SFI to the high voltage output located at the
front of the VERITY 1910 MS Detector (Figure 11).
7. connect the SFI waste tubing (1/32" oD, 0.010" ID PEEK) to the waste line (Figure 12).
8. connect the SFI input tubing (1/32" oD, 0.005" ID PEEK) to the input line (Figure 12).
9. close the door on the front cover.
NOTICE
22
InSTallaTIon
Ensure that the spraychip does not collide with the metalwork.
If there is resistance when tightening the screw, stop and repeat steps 3–5.
Figure 11
The spraychip Assembly Installed on the VERITY® 1910 MS Detector
Figure 12
Waste and Input Line Connections
Ensure the PEEK tubing lines from the SFI are fed through the cutouts in the door.
|
VERITY® 1910 MS DETEcToR
Input
line
Waste line
High
Voltage
connector
Tubing
cutouts