Gilson VERITY 1910 MS User Manual page 14

Hide thumbs Also See for VERITY 1910 MS:
Table of Contents

Advertisement

Specification
Mass Accuracy
Mass Analyzer
Mass Range
Mass Resolution
Nitrogen Gas
Power Requirements
Pump Exhaust
Pumping System
Safety and
Compliance
Sample Inlet
Scan Modes
Scan Rate
Sensitivity
Software Control
Weight
14
InTRoDUcTIon
VERITY® 1910 MS Detector
Description
± m/z 0.3 in full scan*
*In a temperature controlled environment, 20°C ±3°C
ionchip® quadrupole mass filter
m/z 50–1400 with ionchip® 150
m/z 0.7 ± FWHM*
*In a temperature controlled environment, 20°C ±3°C
Pressure
2–6 bar (29–87 psi)
Purity
≥ 99.5%
Site must use either stainless steel or polytetrafluoroethylene (PTFE)
Tubing
tubing.
Stainless Steel 6 mm push lock fitting
Fitting
¼"–6 mm adapter available
Line Voltage
Line Frequency
Power
Overvoltage Category II
Tubing
8 mm OD push lock fitting
Integrated oil-free pumps
The instrument has been certified to safety standards specified for Canada, Europe,
and the United States. Refer to the instrument rear panel label and the Declaration of
Conformity document for the current standards to which the instrument has been found
compliant.
Compatible with standard capillaries and unions e.g., 1/32" tubing
Full scan, SIM (4 m/z channels), simultaneous scan/SIM, and timed SIM
m/z 1,500/s (7,500 pts/s)
10 pg of reserpine yields a peak in SIM mode with a S/N ratio of 10:1 (RMS)
PC control via Ethernet and TRILUTION® LC Software
Control via Ethernet and Gilson Glider Prep (GGP) Software and Masscape® Software
32 kg (70.5 lbs.)
|
VERITY® 1910 MS DETEcToR
100–240V AC ±10%
50/60 Hz
300 W (peak)
For transient overvoltages typically present on the
power supply.
For temporary overvoltages on the power supply.