Plumb the Syringe Pump

1. complete the required fluidic connections and tubing to the syringe. Do not
connect the syringe pump tubing (11) to the microfilter tubing (10) attached to the
split flow interface (SFI). Refer to Figure 18.
connection
11
10
H
G
J
K
2. Flush the syringe pump tubing with calibration solution by starting the pump at a flow rate
between 0.3–2 µL/min until a drop of liquid appears at the free end. continue flushing the
line into a waste container for five minutes.
3. connect the syringe pump tubing (11) to the inline microfilter (G) and
microfilter tubing (10) attached to the SFI.
38
AUTo TUNE (cALIbRATE)
Figure 18
Plumbing Connections between the Syringe Pump
and Split Flow Interface (SFI)
Syringe Pump Tubing
Tubing
1/32" OD, 0.007" ID
PEEK
1/32" OD, 0.005" ID
PEEK
Syringe Pump Components
Syringe
Inline Microfilter
Luer (part number 49041126)
Luer Adapter
(part number 49041125)
|
VERITY® 1910 MS DETEcToR
Fitting
Input
Luer
Luer Fitting
Adapter
Assembly
Coned 6/32
Inline
PEEK
Microfilter
L
M
N
Destination
Inline Microfilter
Split Flow Interface
Luer Fitting (part number 49041125)
6/32 Fitting (part number 49041127)
Plug (part number 49041124)