Remote Monitoring of Output Voltage and Current

Readback
The J2 connector on the rear panel provides access to calibrated readback signals for
Signals
remote monitoring of the output voltage and current. Use rear panel switches SW1-5
and SW1-6 to select either a 0-5 Vdc or a 0-10 Vdc range for the output. The
readback signal represents 0 to 100% of the power supply's output.
See
monitoring of readback signals with 0-5 Vdc or 0-10 Vdc outputs. Use shielded pair
wiring (20 to 24 AWG) and ground the shield to J2 connector terminal 1 or to the
chassis. See
monitoring.
Table 4.2 Power Supply Settings for Remote Monitoring of Readback Signals
Readback Signal
Output Voltage
(0-5 Vdc)
Output Voltage
(0-10 Vdc)
Output Current
(0-5 Vdc)
Output Current
(0-10 Vdc)
Note:Check the readback accuracy any time you reset the SW1 switch. Access offset
and range adjustment for the readback signals through the cover of the unit.
See

Figure 4.3 0-5 Vdc Output Voltage and Current Monitoring

Revision A
Table 4.2
for the required J2 connections and switch settings for remote
Figure 4.3
and
J2 Connection
Signal (+)
J2-9
J2-9
J2-10
J2-10
"Calibrating for Readback Accuracy" on page
Remote Monitoring of Output Voltage and Current
Figure 4.4
for sample configurations using remote
J2 Connection
Return (-)
J2-6
J2-6
J2-6
J2-6
Remote Operation
Switch SW1 Setting
SW1-5 ON (CLOSED)
SW1-5 OFF (OPEN)
SW1-6 ON (CLOSED)
SW1-6 OFF (OPEN)
80.
SW1-5 and SW1-6:
ON (CLOSED)
71