Current Programming Circuit Calibration; Current Programming Calibration Procedures - Xantrex XKW 8-350 Operating Manual

Xkw 3000 watt series programmable dc power supply
Table of Contents

Advertisement

Current
Programming
Circuit
Calibration
Table 4.2 Current Programming Calibration Procedures
Function & designation
of multiturn trim pots
Current Program Offset
(R95)
Current Program Scale
(R66)
Repeat this procedure until the current programming is adjusted to the desired
accuracy.
Notes:
1. The DC shunt-DVM combination must be rated better than 0.5% accuracy. The
2. The required DVM reading is calculated using V=I x R where V is the DVM
Revision A
Step
Action
1
Connect a shunt and DVM to the power supply output. See Note 1.
2
Apply 1% of program voltage. Example: 0.1 for 0-10Vdc or 0.05 for
0-5Vdc programming source
3
Adjust R95 until the DVM indicates 1% of the rated output current.
Example: 0.75A for 40V Model. See Note 2.
1
Connect a shunt and DVM to the power supply output. See Note 1.
2
Apply 100% of program voltage. Example: 10V for 0-10Vdc or
5V for 0-5Vdc programming source
3
Adjust R66 until the DVM indicates 100% of the rated output current.
Example: 75A for 40V Model. See Note 2.
recommended current ratings for the DC shunt and connecting wire must be at
least 10% greater than the power supply's output current rating.
reading, I is the current, and R is the DC shunt resistance.
Calibrating for Programming Accuracy