Remote Monitoring Of Output Voltage And Current; Readback Signals - Xantrex XPR 10-600 Operating Manual

Xpr 6000 watt series programmable dc power supply
Table of Contents

Advertisement

Remote Operation

Remote Monitoring of Output Voltage and Current

Remote Monitoring of Output Voltage and Current
Readback
The J1 connector on the rear panel provides access to calibrated readback signals for
remote monitoring of the output voltage and current. Rear panel switches S1-3 and
Signals
S1-4 allow you to select either a 0-5 Vdc or a 0-10 Vdc range for the output. The
readback signal represents 0 to 100% of the power supply's output.
See
monitoring of readback signals with 0-5 Vdc or 0-10 Vdc outputs. Use shielded pair
wiring (20 to 24 AWG) and ground the shield to the chassis.
Table 4.4 Power Supply Settings for Remote Monitoring of Readback Signals
Readback Signal
Output Voltage
(0-5 Vdc)
Output Voltage
(0-10 Vdc)
Output Current
(0-5 Vdc)
Output Current
(0-10 Vdc)
64
Table 4.4
for the required J1 connections and switch settings for remote
J1 Connection
Signal (+)
J1-4
J1-4
J1-5
J1-5
J1 Connection
Switch S1 Setting
Return (–)
J1-6
S1-3=
J1-6
S1-3=ON
J1-6
S1-4=
J1-6
S1-4=ON
Operating Manual for XPR Series Power Supply
OFF
OFF