Remote Programming of Output Voltage and Current Limit
Remote
Programming
Switch SW1
The output voltage and current limit of the power supply can be remotely
programmed through the rear panel J3 connector
sources, current sources, and resistances. Switch SW1 on the A2 printed circuit
board controls the programming as diagrammed below. When the supply is
controlled by remote programming, the green REMOTE led on the front panel is
illuminated.
Figure 3.6 Switch SW1 Functions
Notes:
•
•
Release 2.2
CAUTION
!
The remote programming input is internally referenced to the supply's negative
output. Do not connect remote programming input lines (J3 pins 9 and 10) to the
supply's positive output.
(Default Settings Shown)
SW-1 and SW-2 are used in combination to select 0-5Vdc, 0-10Vdc, and
0-100mA. See
Table 3.1, "Program Voltage Selection," on page 49
To set switch SW1 shut the unit off, disconnect it from its AC source and remove
the cover. Make the appropriate switch settings then reinstall the cover and
reconnect the unit to its AC source.
Remote Programming of Output Voltage and Current Limit
(Figure
Operation
3.6) using external voltage
for settings.
47