Preparing for storage or transport
WARNING
Risk of tipping over during transport
The medical device may tip over if handled
incorrectly. Observe the following points
when transporting medical devices:
– The medical device may only be moved by
people who have the physical ability to do
so.
– To improve the maneuverability, transport
the device with 2 persons.
– When transporting over inclines, around
corners, or over thresholds (e.g., through
doors or in elevators), make sure that the
medical device does not bump against
anything.
– Remove any devices mounted to the
holding arms or the top of the device.
– Clear the writing tray and fold it down
completely or slide it into the device.
– Do not pull the medical device over hoses,
cables, or other obstacles lying on the
floor.
– Do not activate the brake while the
medical device is being moved.
– Always use the handles on the device to
push or pull it.
1 Press the
key (A) and confirm.
Monitoring and alarm monitoring are switched
off. The ventilator stops.
2 Set the control dial (B) of the vaporizer until it
engages on 0.
Instructions for use Fabius Tiro SW 3.n
3 Close the flow control valves.
4 Close the cylinder valves of the gas cylinders.
5 Pull off the O
and expose to the ambient air.
6 Switch off Fabius with the On/Off switch (B) on
the rear and pull out the plug.
7 Remove hoses of the anesthetic gas receiving
system.
C
A
8 Remove central supply hoses (C).
9 To bring the entire system to normal pressure,
press the
sensor from the inspiratory valve
2
B
280 - 600 kPa
280 - 600 kPa
280 - 600 kPa
O2
AIR
N2O
key.
Operation