Removing The Compact Breathing System; Removing The Compact Breathing; System; Removing The Inspiratory Valve - Dräger Fabius Instructions For Use Manual

Mri
Table of Contents

Advertisement

Removing the Compact Breathing Sys-
tem
Remove all breathing hoses.
Disconnect the breathing bag extension and
bag by loosening the two thumb screws.
Remove the ventilation hose.
Remove the fresh gas hose from the breathing

system.

Remove the scavenger hose.
Remove the flow sensor cable.
Remove the O
sensor cable.
2
Remove the breathing pressure cable.
Detach the APL-bypass and the PEEP/P
lines from the breathing system and from the
side of the machine.
Remove the absorber (for complete instruc-
tions, see page 189
Remove the compact breathing system.
1 Remove plug for the inspiration dome.

Removing the Inspiratory Valve

2 Unscrew the retaining nut.
3 Remove the inspection cap.
4 Extract the valve disc.
5 Extract the gasket on top of the valve disc.

Removing the Expiratory Valve

6 Unscrew the retaining nut.
7 Remove the inspection cap.
8 Extract the valve disc.
9 Remove the gasket on top of the valve disc.
Instructions for Use Fabius MRI SW 3.n
MAX
2
1
3
4
5
Cleaning
6
7
8
9
171