3. Disposal
This product has heavy metal parts. When disposing of this product, comply with local laws and
regulations in your area. Determine whether or not the product is to be treated as a biohazard,
then handle and dispose of accordingly.
4. " Warning" and " Caution" Messages Appearing in Individual Chapters
Chapter 6 Preparation and Inspection
6.1 Connecting PB-30 and an External Monitor
If an external monitor is used, connect the power cable to receptacles for peripherals on the
cart or a receptacle via an insulating transformer. Not doing so may increase the enclosure
leakage current, posing a risk of injury or electric shock to the patient and/or the physician
when he/she comes into contact with those devices.
Connect the external monitor in accordance with the specification of the insulating transformer.
6.3 Installation of PB-30
Use the power source at the rated voltage. Only use the power cord provided with this product.
Improper use of power may cause electric shock or fire.
Connect the power plug to the protective earth receptacle. Do not set up this product in locations
where there is any risk of contact with water. No one except service personnel should open
the cover. There is a risk of accidents by electric shock.
Do not take power from a power strip placed on the floor.
There is a risk of accidents by electric shock.
Do not take power for the second power strip from the first power strip.
Impedance of protective earth may increase and there is a risk of accidents by electric shock.
Do not connect any unit other than the equipment composing this system to the power strip.
Leakage current may increase, posing a risk of injury or electric shock to the patient and/or
the physician when he/she comes into contact with the device.
If the power cords of the peripherals are connected without using the insulation transformer,
the enclosure leakage current may increase, posing a risk of injury or electric shock to the
patient and/or the physician when he/she comes into contact with those devices.
Do not simultaneously touch the patient and any terminals of the devices that make up the
system. Doing so may cause electric shock.
Do not use the equipment in atmosphere of flammable gas. Doing so may cause explosion or fire.
If the inspection result shows any abnormality, do not use the equipment. Use of faulty
equipment may cause misdiagnosis or injury.
Chapter 7 How to Use the Equipment
7.1 Operating the Balloon
Do not operate the balloon forcibly. Doing so may damage mucous membrane.