Chapter 12 Troubleshooting
The balloon on the
over-tube inflates even
though the toggle switch
2 (endoscope side) is
pressed to feed air.
The balloon on the over-
tube does not inflate even
though the toggle switch
on the over-tube side is
pressed to feed air.
The balloon on the
endoscope inflates even
though the toggle switch 1
(over-tube side) is pressed
to feed air.
The balloon dose not
deflate even though the
toggle switch pressed to
evacuate air.
The balloon on the
endoscope is inflated/
deflated slowly.
68
1) The tube 1 (over-tube side) of
TY-04/TY-06 is connected to the
tube connector 2.
2) The plate of TY-400/TY-500 is
installed upside down.
1) The luer lock of the tube
(over-tube side) to be connected
to the over-tube's air feed inlet is
disconnected.
2) The tube (over-tube side) is
connected to the over-tube's
water feed inlet.
3) Broken balloon
4) Incorrect connection of the filter
and tube.
1) The tube 2 (endoscope side) of
TY-04/TY-06 is connected to the
tube connector 1.
2) The plate of TY-400/TY-500 is
installed upside down.
1) Deterioration of pump or break
down of solenoid valve.
2) Faulty filter
1) The tube 2 (endoscope side)
of TY-04/TY-06 is almost
disconnected.
2) The tube (endoscope side)
of TY-400/TY-500 is almost
disconnected.
3) The packing on the tube
(endoscope side) of TY-06/
TY-500 is disconnected or
damaged.
1) Reconnect the tube 1 (over-tube side) of TY-04/
TY-06 to the tube connector 1 and the tube 2
(endoscope side) to the tube connector 2.
2) Install the tube kit so that the letters on the plate
of TY-400/TY-500 are in the correct direction.
1) Connect the luer lock to the balloon air inlet of
the over-tube.
2) Disconnect the tube for the over-tube connected
to the over-tube water inlet and reconnect it to the
air inlet.
3) Replace the over-tube with a new one.
4) Disconnect once and then reconnect.
1) Reconnect the tube 1 (over-tube side) of TY-04/
TY-06 to the tube connector 1 and the tube 2
(endoscope side) to the tube connector 2.
2) Install the tube kit so that the letters on the plate
of TY-400/TY-500 are in the correct direction.
1) Remove the tube from the air inlet of the
endoscope and over-tube and carry out evacua-
tion manually using a syringe. Then, withdraw
the endoscope and over-tube while making sure
of safety on the monitor screen and stop the
examination. Stop the examination and contact
your local FUJIFILM dealer.
2) Replace the filter with a new one.
1) Connect the tube 2 (endoscope side) of TY-04/
TY-06 securely.
2) Connect the tube (endoscope side) of TY-400/
TY-500 securely.
3) Replace the tube kit (TY-06/TY-500) with a new
one.