Chapter 6 Preparation and Inspection
6.2.1 Inspecting the Filter and Tube
(1) Clean and disinfect (or sterilize) the filter and tube of the
tube kit in accordance with Chapter 8 "Cleaning" and
Chapter 10 "Sterilization."
(2) Make sure that there is no irregularity such as a flaw or
deformation in the appearance of the filter and tube.
[Note]
If any abnormality such as a flaw or deformation is found,
replace it with a new one.
For TY-06 and TY-500, make sure that there is no
irregularity such as a flaw or deformation in the
appearance of the packing on the connector of the tube
(endoscope side).
[Note]
The filter and tube are consumable supplies. Replace
them with new ones once a month or once every 10 cases,
whichever comes first.
[Note]
If body fluid flows back to the filter and is trapped there,
be sure to replace the complete set of the filter and tube
kit.
[Note]
Be sure to turn off the power switch before disconnecting
the filter and tube from the main unit.