Electromagnetic immunity compliance information and guidance
Immunity test
Conducted RF
3Vrms
EN 61000-4-6
150kHz to 80MHz
3V/m
Radiated RF
80MHz to 2.5GHz
EN 61000-4-3
Electromagnetic immunity compliance information and guidance
The customer or the user of this product can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitter) and this product as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter P (W)
0.01
0.1
1
10
100
Compliance
EN 60601-1-2
Test level
level
3V[V
3V/m[E
Separation distance related to frequency of the transmitter (m)
150kHz to 80MHz
d=1.2
√ P
0.12
0.38
1.2
3.8
12
Portable and mobile RF communications equipment
should be used no closer to any part of this product,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
3.5
d=
√ P
V
1
3.5
d=
√ P
E
1
7
d=
√ P
E
]
1
1
]
Where "P" is the maximum output power rating of the
1
transmitter in watts (W) according to the transmitter
manufacturer and "d" is the recommended separation
distance in meters (m).
This product complies with the requirements of
EN 60601-1-2: 2007.
However electromagnetic interference may occur on this
product under electromagnetic environment that exceeds
its noise level.
Electromagnetic interference may occur in the vicinity
of equipment marked with the following symbol.
80 to 800MHz
d=1.2
√ P
0.12
0.38
1.2
3.8
12
Guidance
80 to 800MHz
800MHz to 2.5GHz
800MHz to 2.5GHz
d=2.3
0.23
0.73
Main Specifications
√ P
2.3
7.3
23
77

Hide quick links: