Chapter 9 Chemical Disinfection
9.1 Methods of Chemical Disinfection
Prior to disinfection, cleaning should be performed according to Section 8.3.
Table 9.1 Compatible methods of cleaning, disinfection and sterilization for PB-30 and consumable supplies
Main unit
Remote switch
Tube
Tube kit
Filter
Balloon
[Note 1] Follow the conditions of disinfection (Section 9.2).
[Note 2] Follow the sterilization procedure for the balloon (Section 10.4).
Method of cleaning
Chemical
Ultrasonic
cleaning
cleaning
No
No
No
No
Yes
Yes
No
No
Yes
No
Chapter 9 Chemical Disinfection
Method of disinfection and sterilization
Chemical
Autoclaving
disinfection
No
No
No
No
Yes
Yes
Yes
No
[Note 1]
No
Yes
Gas
sterilization
No
No
No
No
[Note 2]
Yes
55