Chapter 9 Chemical Disinfection
(8) Keep the balloon immersed in the chemical solution in
accordance with the temperature, concentration and time
recommended by the chemical solution manufacturer, and
cover the basin to prevent the vaporization of the chemical
solution.
[Note]
Confirm that no potions of the balloon fold over onto itself
which may prevent contact of balloon surfaces with the
chemical solution.
(9) Remove the balloon from the chemical solution.
(10) Fill the basin with sterile water.
(11) Soak the balloon in sterile water, and rinse it well while
gently shaking.
(12) Clean the inside of the balloon by kneading it with fingers
for at least 15 seconds.
(13) Flush sterile water with a syringe into all openings,
internal areas and gaps of the balloon with a syringe, and
wipe its surfaces with sterile gauze in sterile water to
remove air bubbles completely.
(14) Remove the balloon from sterile water.
(15) Wipe off the remaining water with sterile gauze.
(16) Dry the balloon thoroughly.