9.4 Disinfection of Tube Kit

(1) Fill a basin with chemical solution prepared according
to the chemical solution manufacturer's instructions to
completely immerse the entire tube kit.
(2) Completely immerse the tube kit in the chemical solution.
(3) Inject 30 mL or more of chemical solution into the tube kit
until air bubbles disappear.
(4) Inject chemical solution into all openings, internal areas
and gaps of the tube kit with a syringe, and wipe the
surface of it with gauze to remove all air bubbles.
(5) Keep the tube kit immersed in the chemical solution for
the contact conditions recommended by the chemical
solution manufacturer, and cover the basin to prevent the
vaporization of chemical solution.
[Note]
When carrying out disinfection, follow the conditions
indicated in "9.2 Conditions of Disinfection."
(6) Pull the tube kit out of the chemical solution.
(7) Inject air with a syringe into the tube kit to discharge the
chemical solution.
(8) Fill sterile water in the basin which can immerse the tube
kit completely.
The tube kit must be completely immersed in a chemical solution.
Remove air bubbles completely. If any air bubbles remain, effective
disinfection cannot be achieved and an inadequately disinfected tube
kit may pose an infection risk.
WARNING
Chapter 9 Chemical Disinfection
59