8.3 Cleaning

8.3.1
Cleaning Tube Kit
(1) Fill a basin with fresh cleaning fluid prepared according
to the cleaning fluid manufacturer's recommendations to
completely immerse the tube kit.
(2) Immerse the entire tube kit in the cleaning fluid.
(3) Clean the surface of the tube kit in the cleaning fluid using
a clean lint-free cloth.
(4) Attach a syringe filled with the cleaning fluid to the
balloon controller side connector. Flush 30 mL or more
of cleaning fluid into the tube kit with a syringe until air
bubbles disappear.
(5) Inject the cleaning fluid with a syringe into all openings,
internal areas and gaps of the tube kit in the cleaning fluid
to remove all air bubbles.
(6) Keep the tube kit immersed in the cleaning fluid for
the contact conditions recommended by the cleaning
fluid manufacturer and cover the basin to prevent the
vaporization of cleaning fluid.
Chapter 8 Cleaning
53