Chapter 10 Sterilization

10.4 Sterilization of Balloon

Table 10.2 Conditions of EOG sterilization (chamber type)
Concentration of
ethylene oxide
20 v/v%
(1) Before gas sterilization, clean the balloon, rinse it with
water, and remove moisture from all balloon surfaces.
Insufficient removal of moisture may compromise
sterilization.
(2) Gas-sterilize it according to the conditions shown in Table
10.2. After sterilization, aerate it thoroughly.
64
Proceed with gas sterilization after removing moisture from the balloon.
Applying gas sterilization to wet surfaces may compromise sterilization.
Proceed with aeration after gas sterilization. Gas residuals remaining
in the balloon after gas sterilization may be harmful to the human body.
Use gas sterilization to sterilize the balloon.
For gas sterilization, the specific parameters described in Table 10.2 are recommended.
We have confirmed that our equipment is compatible with the sterilization conditions
shown below.
Temperature
55ºC
WARNING
CAUTION
Humidity
Pressure
Normal
100 kPa or more
humidity
(relative pressure)
Time
Aeration
Atmospheric
4 hours
pressure
Temperature
Time
55ºC
12 hours